Imagine an ideal 3 V source feeding two equal 1 MΩ resistors in series. Without the meter, the midpoint is at 1.500 V relative to the lower end. A 10 MΩ meter across the lower resistor makes that lower branch equal to 0.909 MΩ. The upper resistor remains 1 MΩ. The indicated voltage is then 3 × 0.909 / (1 + 0.909), or approximately 1.429 V. The reading is about 4.76% below the unloaded prediction. No broken component is needed to explain the difference.

Change the circuit scale, not the algebra

Repeat the same calculation using two 330 Ω resistors. The lower branch becomes 329.989 Ω and the midpoint becomes approximately 1.499975 V. The change is about 25 microvolts, written 25 µV. An inexpensive handheld meter may not resolve it. The calculation does not authorize reporting those digits as an observed result. It explains why a low-resistance teaching circuit and a high-resistance sensor circuit can react differently to the same instrument.

United States · instrument records

A source has its own contribution

An ideal source maintains its stated voltage regardless of load. A simple real-source model adds a series resistance. If its open-circuit voltage is 3.00 V and its modelled internal resistance is 10 Ω, a 330 Ω load draws 3 / 340 A, about 8.82 mA. The load receives about 2.912 V. The 10 Ω value is a chosen example, not a measurement of the AA cells in your holder. Battery chemistry, condition, temperature and time make an actual cell more complicated than one fixed resistor.

Separate loading from ordinary drift

A declining reading could come from a changing source, a changing load, a warming component, a poor connection or the meter itself. Do not assign it to one cause because that cause is interesting. Keep the load fixed, return to the same probe points, note elapsed time and repeat the open-circuit and loaded readings in a documented order. If you estimate source resistance from voltage change divided by load current, identify it as an operating-point estimate. It is not a universal battery specification.

An instrument manual is part of the apparatus

The Fluke manuals cited here describe measurement effects that a simple circuit icon cannot show. The transferable lesson is to inspect the input behaviour and range specification of the actual instrument. A different meter may change resistance with range or use additional input circuitry. On a learning bench, the honest conclusion may be that the predicted loading is smaller than you can resolve. That is a result about the experiment’s reach, not a reason to invent extra decimal places. [R04]

United States · leads, contacts and sense points

Units and terms in this account

V volt, electrical potential difference · % percent, parts per hundred · = equals · ≈ approximately equal to · × multiplied by

Ω ohm, electrical resistance · V volt, electrical potential difference · AA A standardized cylindrical battery size designation, not an acronym · A ampere, electric current · mA milliampere, 0.001 ampere

Ω ohm, electrical resistance · = equals · % percent, parts per hundred