Separate loading from ordinary drift
A declining reading could come from a changing source, a changing load, a warming component, a poor connection or the meter itself. Do not assign it to one cause because that cause is interesting. Keep the load fixed, return to the same probe points, note elapsed time and repeat the open-circuit and loaded readings in a documented order. If you estimate source resistance from voltage change divided by load current, identify it as an operating-point estimate. It is not a universal battery specification.
United States · instrument records
An instrument manual is part of the apparatus
The Fluke manuals cited here describe measurement effects that a simple circuit icon cannot show. The transferable lesson is to inspect the input behaviour and range specification of the actual instrument. A different meter may change resistance with range or use additional input circuitry. On a learning bench, the honest conclusion may be that the predicted loading is smaller than you can resolve. That is a result about the experiment’s reach, not a reason to invent extra decimal places. [R04]
United States · leads, contacts and sense points
Measure the object, or measure its connections?
Four wires place the boundary at the device
The sense leads do not carry the main test current in this model
TWO-WIRE: the leads enter the result
Read the illustration’s labels
Test source
and reading
0.20 Ω
0.10 Ω
Indicated total: 0.20 + 0.10 + 0.20 = 0.50 Ω
FOUR-WIRE: define the sense contacts
Current source
known I
Device
S+
S−
V
Separate voltage-sense pair
R = sensed V / test I · the chosen contact positions define the object
Original teaching diagram · stated examples, not recorded measurements
The resistance that travelled with the probes
In a two-wire resistance measurement, the instrument’s test current passes through its leads, contacts and the device. A reading therefore includes more than the intended component. With a 0.10 Ω object and 0.20 Ω in each lead-and-contact path, the simple indicated total is 0.50 Ω. Calling all of it the object’s resistance creates a 400% error relative to 0.10 Ω. For a 330 Ω resistor the same extra 0.40 Ω is only about 0.12%. The importance depends on scale. [R19]
Units and terms in this account
Ω ohm, electrical resistance · V volt, electrical potential difference · AA A standardized cylindrical battery size designation, not an acronym · A ampere, electric current · mA milliampere, 0.001 ampere
Ω ohm, electrical resistance · = equals · % percent, parts per hundred