A four-wire or Kelvin arrangement uses one pair of leads to supply current and a separate pair to sense voltage at the chosen boundary of the object. Because the sense input draws very little current, its lead drops can be small. Dividing the sensed voltage by the applied current gives a result that excludes the current-lead drops outside those sense points. The drawing makes the boundary visible. It does not turn a two-terminal handheld meter into a four-wire instrument. [R05]

Where the sense points go matters

If a sense contact sits on the far side of an oxidized joint, that joint may remain inside the measured boundary. Move it closer to the material and you may be asking a different question. Neither reading is automatically wrong. One might describe a complete connector assembly; the other might describe a metal coupon. Label the points on a drawing so a later reader knows which resistance was included.

United States · leads, contacts and sense points

Why subtracting a short is limited

Shorting the probes together and recording their reading can reveal a substantial lead contribution. Subtracting that value assumes the contribution remains stable when the probes are moved to the object. Contact pressure, surface condition and temperature may change. A zero or relative function is therefore a convenience with assumptions, not a guarantee of removed uncertainty. On our 330 Ω teaching bench, record the shorted-lead check as diagnostic information rather than advertising a precision result the instrument cannot support.

Reversal can expose a concealed voltage

A tiny thermoelectric voltage can arise where dissimilar conductors meet at different temperatures. In a suitable laboratory system, taking readings at equal positive and negative test currents can help separate a resistance-related voltage, which reverses, from an approximately constant offset, which does not. For example, +10 mA gives +1.02 mV and -10 mA gives -0.98 mV. The difference divided by 20 mA gives 0.100 Ω; their average is +0.02 mV. These are constructed numbers. They assume the object and offset remain stable between readings. [R03]

Units and terms in this account

Ω ohm, electrical resistance · mA milliampere, 0.001 ampere