Read the illustration’s labels
Series-parallel impedance: magnitude and phase
Model: Rₛ = 1 kΩ; Rₚ = 10 kΩ; C = 10 µF
0.01
0.1
1
10
100
1,000
3,000
10,000
|Z| (Ω)
0°
-30°
-60°
-90°
Phase
Phase minimum: 5.279 Hz
Frequency (Hz), logarithmic scale
Calculated ideal model. Its phase minimum is not the Nyquist semicircle apex.
For Rs > 0, fphase = √(1 + Rp/Rs) / (2πRpC). A fitted circuit still needs physical evidence.
Electrochemical impedance spectroscopy (EIS) measures the response to a small alternating perturbation over frequency. Impedance Z is a complex ratio of voltage to current. Its magnitude has units of ohms but is not generally identical to resistance. The figures here use one explicitly ideal electrical circuit.
Use a series resistance Rs followed by a parallel resistance Rp and capacitance C. Its model is Z = Rs + Rp/(1 + jωRpC), where j² = -1 and angular frequency ω = 2πf. In a Nyquist plot with Z′ horizontally and -Z″ vertically, the ideal curve runs between Rs and Rs + Rp. The semicircle apex has coordinates Rs + Rp/2 and Rp/2. Frequency labels are essential because frequency is not one of those two axes.
Take Rs = 1 kΩ, Rp = 10 kΩ and C = 10 µF. The apex frequency is 1/(2πRpC), approximately 1.592 Hz. A Bode plot shows magnitude and phase against frequency. Its phase peak is a different feature when Rs is nonzero: for this model it occurs at sqrt(1 + Rp/Rs)/(2πRpC), approximately 5.279 Hz. Substituting that phase-peak frequency into the semicircle-apex formula gives the wrong capacitance.
The distinction follows by writing x = ωRpC. The ratio determining the magnitude of the phase tangent is Rp x/[Rs(1 + x²) + Rp]. Differentiating with respect to x gives a maximum at x² = 1 + Rp/Rs when Rs is positive. The Nyquist apex instead comes from maximising Rp x/(1 + x²), which occurs at x = 1. The two maxima answer different geometric questions.
A fitted circuit is an interpretation, not a photograph of a material. A limited frequency interval can allow several models to fit similar data. A depressed arc is often represented with a constant-phase element, CPE, whose admittance is Q(jω)^α. Q has the dimensions of capacitance only when α = 1. A 45-degree tail can be consistent with a diffusion model under its assumptions, but the shape alone does not prove one unique transport mechanism.
Instrument guidance also requires attention to linearity, stability, causality and stationarity. A corroding or changing sample can evolve during a slow frequency sweep. Adding more frequency points does not repair that change. The circuit model describes the electrical response under stated assumptions. Electrode labels and material behaviour require their own physical evidence.
Source: BioLogic / Checking non‐stationary impedance measurements, application note 69
Z: Complex impedance. Its real and imaginary parts describe the stated frequency response.
Rs: Series resistance in the illustrated impedance model.
| Key | Small-print annotation |
|---|---|
| A | IUPAC / source |
A microscope is a chain of light, support and recording
Contemporary explanatory reconstruction, not a replica fabrication drawing.
Read the illustration’s labels
Optical observation station
The handover must retain
Geometry
Illumination
Observation
Interpretation
Provenance
A
B
C
D
E
F
| Key | Small-print annotation |
|---|---|
| A | Eyepiece |
| B | Reference plane, support and lens position |
| C | Direction, contrast and preparation |
| D | Specimen support |
| E | What is seen; what is inferred |
| F | Maker, plate, date and later reconstruction |
Compare the historic instrument, flea, needle and cork plates with the modern measurement record.
Hooke’s 1665 engravings are historical evidence. This vector explains the observation chain at a schematic scale.
Units and terms in this account
kΩ kilohm, 1,000 ohms · ° degree, an angle unit · Hz hertz, cycles per second · Ω ohm, electrical resistance · = equals · EIS Electrochemical impedance spectroscopy: measurement of electrical response as a function of frequency
kΩ kilohm, 1,000 ohms · Hz hertz, cycles per second · = equals · CPE Constant-phase element, an empirical equivalent-circuit element with a frequency-independent phase angle under its stated model · IUPAC International Union of Pure and Applied Chemistry