Uncertainty is not an apology attached to a number. It describes the dispersion reasonably associated with a result under a stated model. Begin with the equation and identify its inputs. For a photographed shape, use d = (L/p) square root of (4A/π), where L is a reference length, p its span in pixels and A the selected object's area in square pixels. The output d uses the same length unit as L.

For small independent contributions, the relative combined standard uncertainty is the square root of [(uL/L)² + (up/p)² + (uA/(2A))²]. The area term has a factor of one half because diameter depends on the square root of area. Substitution gives approximately 1.04 percent, or 0.0104 mm. A factor of two would

produce an expanded uncertainty of approximately 0.0208 mm under the stated assumptions; it is not automatically an exact 95 percent probability statement.

Check those assumptions before using the formula. If the same scale calibration is applied to every particle, its uncertainty is shared. It does not shrink as the square root of the number of particles merely because more objects are counted. Pixel-edge decisions can also be correlated. A common perspective distortion is not an independent random error on each particle. List unresolved effects rather than hiding them inside an unjustified numerical total.

Separate a standard uncertainty from a tolerance limit, an observed range and an instrument's last digit. They are not interchangeable. Converting a stated bound into a standard uncertainty requires an assumed probability distribution and a reason for using it. If that basis is unavailable, report the bound or observed range as such. A clear limitation is more useful than a formal-looking uncertainty copied from a template.

Open the complete related illustration · page 2113

Wider world

Build a measurement library that survives a broken instrument

Read the illustration’s labels

DESIGN / M-004

BUILD / revision B

OBSERVATION / trial 03

INTERPRETATION / v1

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

P

KeySmall-print annotation
AQuantity: equivalent diameter
BBase: 450 × 350 × 18 mm
CReference plane: board face
DPhone: 248 g; clamp checked
ETarget range: 1 to 20 mm
FReference bar: 100 mm
GLimit: projected area only
HChanged: bolt spacing washers
IOriginal: M004_03.jpg
JEquation: d = s √(4A/π)
K10 mm span: 1000 px
LRaw file remains untouched
MAccepted object IDs: 26
NUnits: millimetres
OExcluded IDs: 6; listed
PUnresolved: perspective bound

A resilient workshop keeps more than a shelf of instruments. It keeps reference objects, spare consumables, written methods, raw records and examples of known failures. Give each reusable setup an identifier and a drawing revision. A copy stand, ruler, lens, capacitor and meter can then be replaced individually without losing the meaning of earlier observations.

Store four kinds of record separately but link them by identifier. The design record states what the apparatus was intended to do. The build record lists the actual materials, dimensions and substitutions. The observation record contains unaltered readings and images. The interpretation record contains calculations, exclusions and conclusions. This separation lets a future reader recompute an answer after discovering a calibration error.

Four linked records make the result recoverable

Use one setup identifier and drawing revision across every sheet.

A new reader should reproduce one result using the record alone, then identify the conditions that would invalidate it.

Units and terms in this account

mm millimetre, 0.001 metre · % percent, parts per hundred · = equals · ≈ approximately equal to

mm millimetre, 0.001 metre

mm millimetre, 0.001 metre · g gram, 0.001 kilogram · = equals · × multiplied by · ARS Argentine peso · m metre, a length unit